Spiral Laminar Flow: The Evidence

Clinical Abstracts relating to Spiral Flow Grafts and Spiral Laminar Flow Technology

Click on one of the topics below to jump to that section

Section A: AV Access Outcome Studies

Section B: Peripheral Bypass Outcome Studies

Section C: In Vivo Human Studies

Section D: In Vitro, Animal and Scientific Studies

Section A: AV Access Outcome Studies

1. Two year follow-up results for Spiral Flow Grafts in AV Access

Presented at the 37th Charing Cross International Symposium, April 2015, London, UK Prim. Univ.-Doz.
Dr. Wolfgang J. Hofmann, Department of Vascular Surgery, Landeskrankenhaus, (LKH) Feldkirch, Austria

Introduction

The AV access strategy at LKH Feldkirch focusses primarily on utilising native fistula and Duplex mapping of the upper extremity is undertaken to identify the most appropriate approach.
AV grafts are only used in patients not suitable for a native fistula.
Conventional AV grafts typically require one to two revisions per year.

Methods

Since September 2010, a series of 16 grafts have been implanted in 15 patients. All patients have had previous surgery (mean of 4.3 previous procedures (0-24).
Mostly loop grafts in the forearm are utilised.

  • 11 loop forearm
  • 2 loop upper arm
  • 2 straight upper arm
  • 1 loop thigh
  • 4 graft occlusions / two successful thrombectomies
  • 4 graft explants (2 due to steal, 2 due to infection)

Results

The results from 16 Spiral Flow™ AV Grafts, implanted between September 2010 and August 2013 were reviewed. Median follow up was 22 months and all grafts were Duplex scanned every three months to confirm the presence of Spiral Flow. Additional scans were performed if there was any suspicion of graft failure.
At 22 months, the primary rate was 72%; secondary patency was 85.5%
The results were compared to a series of 79 conventional PTFE grafts implanted by Dr Hofmann between January 2005 and December 2009. A Kaplan-Meier analysis highlighted primary patency rates at 18 months of 72% for Spiral Flow™ and 36.7% for conventional PTFE grafts. (P=0.01)

Conclusion

Changing the flow pattern at the venous anastomosis of AV grafts using the Spiral Flow Technology seems to be a simple but valuable tool in order to improve patency rates of PTFE AV grafts

Back to top

2. Intermediate results of using spiral flow av graft: is it a breakthrough solution to a difficult problem?

In Abstracts from VAS 9th International Congress, April 15-18, 2015 Barcelona, Spain. J Vasc Access 2015; 16(2): 31
Hosam F. El Sayed, Houston Methodist Hospital, Houston, United States

Introduction

AV access grafts are used in those patients where there are no available superficial veins for native AV fistula creation.
Their usable life and patency rates are far from being ideal requiring frequent interventions to maintain their use for dialysis. Their failure is usually related to neointimal hyperplasia leading to stenosis of the venous outflow near the venous anastomosis. Studies have shown that, Spiral laminar flow is the normal pattern of flow in most of the large and medium sized vessels in vivo as well as many functioning native AV fistulas. The Spiral flow graft has a design that creates a spiral laminar flow at the venous end that is a hypothetically reduces intimal hyperplasia and graft failure. We here report the midterm results of the largest reported series of using the graft in AV access.

Material & Methods

After IRB approval, a prospective study of using Spiral Flow graft for AV access in our institution between Jan 2012 to Dec 2014 was performed. Enrolled patients had no suitable superficial veins for native AV fistula creation. Patient demographics and comorbidities were recorded.
Kaplan Meier curve analysis was used to calculate patency rates compared to historic controls of straight ePTFE and heparin bonded grafts in our institution.
Complications were also recorded.

Results

48 cases were included. The access site was the upper arm (32), the forearm (13), and chest wall (3). Mean age of 61 and mean follow up of 14 months. At 12 month, the assisted primary and secondary patency rates were 70% and 82%, respectively that was significantly better than historic controls using straight ePTFE and heparin bonded grafts in our own institution.
Complications included 4 graft infections; 3 severe steal syndrome, 4 seromas and 3 arm swelling. There were only 2 early graft failures.

Conclusions

Spiral flow grafts are a valid successful option for AV access.
One year results are superior to using straight ePTFE and heparin bonded grafts. This may be explained on the basis of the hemodynamic environment created by the spiral laminar flow and may be a significant contribution to preventing neointimal hyperplasia and hence AV access graft failure.

Back to top

3. Spiral Flow Technology a year on: Results of the North American experience

Presented at the 36th Charing Cross International Symposium April 2014, London, UK
Hosam El Sayed, MD, Assistant Professor of Cardiovascular Surgery, Methodist DeBakey Heart and Vascular Center, Houston, Texas, USA.

Introduction

Prosthetic graft failure is a normal tissue response to an abnormal flow environment. Endothelial cells at the anastomosis are sensitive to non-laminar flow environment (turbulence, stagnation, low shear stress, increased oscillatory index). These cells respond by signalling neo-intimal hyperplasia thus promoting failure. Results with prosthetic AV access grafts are far from ideal. We decided to use the Spiral flow graft for AV access based on early encouraging reports.

Patients and methods

This is a retrospective review of all cases that had Spiral Flow graft placement for AV access in the upper extremities.
From January, 2012 – January, 2014, 38 Spiral Flow™ AV Access Grafts (Vascular Flow Technologies Ltd) were placed in 37 patients.
The mean age was 61 years (range 42-88 years); 47% (18) were male. Patients presented with the following comorbidities:
Diabetes: 25 (66%); Hypertension: 33 (87%); CAD: 8 (21%); CHF: 7 (18%); CVA: 7 (18%); PVD: 8 (21%).
Grafts were implanted as: Upper arm: 24; Forearm: 11; Chest wall: 3

Results

The mean follow-up was nine months. 15 patients completed 12 months follow up. There are 20 grafts (53%) currently in use. Three grafts were removed for infection, 3 grafts were ligated for severe steal and 1 graft was ligated for severe arm swelling. Seven patients are deceased, all with their grafts patent.
Complications were as follows: Graft infection: 3 (8%); Significant Steal: 3 (8%); Thrombosis: 7 (18%); Venous stenosis: 3 (8%); Seroma: 4 (10%); Wound complications: 3 (7%); Venous hypertension: 7(18) and Pseudo-aneurysm: 0
Overall patency (%) results at 12 months are as follows:

Primary AssistedSecondary
81% ± 983% ± 9

Conclusions

Spiral Flow Grafts are a valid and successful option for AV access. Early results are encouraging and tend to be better compared to standard straight ePTFE and heparin bonded grafts.
This may be explained on the basis of improved hemodynamics created by the spiral laminar flow.

Back to top

4. Early results of using Spiral Flow AV Graft: Is it a breakthrough solution to a difficult problem?

Poster at Society for Clinical Vascular Surgery (SCVS) symposium March 2014, Carlsbad, California, USA.
Hosam F. El Sayed, MD, Javier E. Anaya-Ayala, MD, Mark Davies, MD. Houston Methodist DeBakey Heart & Vascular Center, Houston, Texas, USA

Objective

Although, the preferred method to create an access is native AV fistula, there is still a significant number of patients where this is not feasible. AV access grafts are used in those patients and their patency rates are far from being ideal requiring frequent interventions to maintain their use. Their failure is usually related to stenosis of the venous outflow due to intimal hyperplasia, near the venous anastomosis of the graft. Neo-intimal hyperplasia may, in part, be a normal cellular response to an abnormal (turbulent) flow environment created by the AV access. Studies have shown that, Spiral laminar flow is the normal pattern of flow in most of the large and medium sized vessels in vivo. The Spiral flow graft has a design that creates a more natural spiral laminar flow at the venous end that is a hypothetically a more friendly hemodynamic environment thus reducing intimal hyperplasia and graft failure. We here report the early results of the largest available series of using the graft in AV access.

Method

Retrospective review of all cases using the Spiral Flow graft for AV access in our institution, Jan 2012 to Jan 2014. Patients selected had no suitable superficial veins for native AV fistula. Patient demographics and comorbidities were recorded. Kaplan Meier curve analysis was used to calculate primary, assisted primary and secondary patency rates in comparison to historic controls of straight ePTFE and heparin bonded grafts for the same indication in our institution. Complications were also recorded.

Results

A total of 38 cases were included. The access site was the arm (24), the forearm (11), and chest wall (3). Two thirds were females, mean age of 60 years and mean follow up of 8 months. At 12 month, the primary, assisted primary and secondary patency rates were 73%, 73% and 79%, respectively. Complications included 4 graft infections; 3 severe steal syndrome, 4 seromas and 2 arm swelling. There was only 1 early graft failure. There was only one early graft failure in less than 30 days in a patient with known hypercoagulable state who was not therapeutic on anticoagulation.

PrimaryPrimary AssistedSecondary
Spiral AVG73 ± 873 ± 879 ± 8
Heparin bonded56 ± 860 ± 859 ± 8
Standard ePTFE38 ± 1047 ± 10 66 ± 11
P-Value0.0550.160.042

12 month patency rates

Conclusion

Spiral flow grafts are a valid and successful option for AV access. One year results tend to be significantly superior to using straight ePTFE and heparin bonded grafts. This may be explained on the basis of the hemodynamic environment created by the spiral laminar flow and may be a significant contribution to preventing neointimal hyperplasia and hence AV access graft failure.

Back to top

5. Early experience with the SLF (Spiral Laminar Flow) AV Access Graft

In Abstracts from VAS 8th International Congress, April 25-27, 2013 Prague, Czech Republic. J Vasc Access 2013; 14(1): 43
Inston N1, Hofmann WJ2.
1 Department of Renal Surgery, Queen Elizabeth Hospital, University Hospitals Birmingham, Birmingham, UK
2 Department for Vascular Surgery, Feldkirch General Hospital, Feldkirch, Austria

Background

Arteriovenous grafts for dialysis are associated with a high incidence of outflow stenosis requiring re-intervention and increased risk of failure. Pathological flow patterns promote neointimal hyperplasia in the outflow vein and attempts to improve graft outcomes have resulted in innovative graft technologies. One such strategy involves the generation of spiral laminar outflow from the graft and the SLF AV graft (Vascular Flow Technologies, Dundee) is designed to recreate normal patterns of blood flow to reduce outflow stenosis. The aims of this study were to audit the early clinical outcomes of SLF grafts in a patient series from two European units.

Materials & Methods

A prospective series of implants was been collated from two surgeons series. The decision to use an SLF was based on surgeon preference and the site of implant based on clinical parameters. Flow was determined using Doppler ultrasound. For clinical outcomes data was collected prospectively with a standardized data collection tool at set time points post surgery.

Results

Since September 2010 twelve SLF-AV Access grafts were implanted. All grafts were used in patients with complex access histories (median of 6 previous AV access procedures). One graft failed immediately, the cause attributed to poor inflow (small diseased high bifurcation brachial artery and hypotension). Of the eleven with immediate function two failed (56 days and 58 days). Thrombectomy was performed and no outflow stenosis was demonstrated in either. Despite this both grafts re-clotted within 24 hours. Primary patency in the series is 75% at a mean time point of 188 days (range 0 to 448). One superficial wound infection occurred which required surgical intervention. No graft infections occurred requiring graft intervention.

Conclusions

Whilst this represents a small series with limited follow up this early experience of the SLF AV is encouraging. Doppler studies support the SLF generating spiral flow at the distal (graft to vein) outflow and the short term data herein supports a low incidence of outflow stenosis with excellent primary patency to date.

Back to top

6. A new choice for vascular access: Spiral Flow AV Graft

In Abstracts from VAS 8th International Congress, April 25-27, 2013 Prague, Czech Republic. J Vasc Access 2013; 14(1): 36-37
Cetingok U, Department of Cardiovascular Surgery, Kavaklidere Umut Hospital, Ankara, Turkey

Background

The most commonly used grafts for vascular access are the polytetraflouroethylene (PTFE) grafts. Different types of PTFE grafts are used in the vascular access applications for a long time. Neointimal hyperplasia at the distal anastomosis is one of the primary failure modes for prosthetic grafts resulting in occlusion and loss of function. The Spiral Flow Vascular Access Graft is a wrapped, expanded polytetrafluoroethylene (ePTFE) graft with custom features injection moulded of Chronoflex polyurethane. The distal portion of the graft contains the patented Spiral Flow™ Inducer, the Inducer Indicator Ring and pre-cut distal anastomotic cuff. The injection moulded components of the graft are non-removable.

Materials & Methods

From July 2010 until January 2013, Spiral Laminar Flow PTFE Graft (Vascular Flow Technologies Ltd., UK) was implanted in 16 patients for vascular access. Primary graft implantation was performed in 2 patients.

Results

Mean follow-up was 26 months. The complications as graft infection, pseudoaneurysm, venous hypertension was not seen and thrombectomy was performed two patients due to graft thrombosis. Primary patency rate was 88% and secondary patency rate was 100% during follow-up.

Conclusions

The process of blockage caused by neointimal hyperplasia in and near the graft-vein anastomosis is driven to a large part by disturbed or turbulent flow. The Spiral Flow™ Vascular Access Graft delivering less turbulent energy by utilizing Spiral Laminar Flow™ delivering blood flow into the venous system like a native AV fistula should: Improve device patency, prolong the effective life of devices, reduce downstream disease progression, improve patient quality of life. This flow pattern can be determined with color doppler ultrasound study using a transverse interrogation of the vessel at low velocity settings. The “red/blue split” seen in transverse color doppler ultrasound study is characteristic of spiral laminar flow and is usually seen in healthy arteries. The goal of this study is to determine if the SLF™ Graft creates spiral laminar flow and if that might contribute to improved patency.

Back to top

7. Vascular Flow Technology: Another run of the mill graft or a breakthrough technology?

Experience and perspective from a European centre.
Presented at the 35th Charing Cross International Symposium, 7th April, 2013, London, UK
Prim. Univ.-Doz. Dr. Wolfgang J. Hofmann
Department of Vascular Surgery, Landeskrankenhaus (LKH) Feldkirch, Austria

Background

The AV access strategy at LKH Feldkirch focusses primarily on utilising native fistula and Duplex mapping of the upper extremity is undertaken to identify the most appropriate of the following approaches:

  • Cimino Brescia Fistula
  • Cubital Fistula
  • Ulnar Artery – Basilic Vein Shunt
  • Transposition and Shunting of the Basilic Vein

AV prosthetic grafts are only used in patients if using a native fistula is not possible. So these are negatively selected. Conventional AV grafts typically require one to two revisions per year.
There is neointimal hyperplasia at the heel, toe and bed of the anastomosis. This is due to turbulence at the venous anastomosis.
In order to prevent stenosis and occlusion a number of tactics were tired unsuccessfully in the last 10 years, including:

  • Configuration of the Anastomosis (Venaflo®)
  • External beam radiation
  • Patchplasty
  • PTA (repetitive)

On introduction to the Spiral Flow graft in 2010, a Duplex scan confirmed that this spiral movement in a normal healthy artery is real and if you can see it. The first Spiral Flow graft implantation was undertaken in September 2010.
More than 26 months after implantation, that graft is working well and we have spiral flow at the venous anastomosis.

Method

Since September 2010, a series of 10 cases have been completed so far. All patients have had previous surgery (mean of 3.8 previous procedures; median of 3) Only three patients were diabetic, which is less than our average. Mostly loop grafts in the forearm are utilised.

Results

Only one problem was observed. One patient had an early occlusion of graft only two months after implantation. A revision was undertaken thinking that the venous anastomosis could be slightly better in a more proximal position. Another spiral graft was implanted but it thrombosed also. As it was discovered that this patient is an alcoholic, who is said to sleep with his head on his arm (directly over the graft), it was decided to abandon further graft therapy (the patient is now on a dialysis cannula).
Currently, 10 grafts have been implanted in nine patients with a mean follow-up of six months.
Only two failures were observed, both in the same patient. All other grafts are patent and on haemodialysis with no evidence of neointimal hyperplasia (NIH) under Duplex review.

  • 10 grafts in nine patients
  • Mean follow up 6.1 months (range: 31 – 1 months)
  • Two early failures in one patient (patient deemed not suitable for treatment)
  • All other grafts patent (80%) without evidence of NIH in Duplex control

Conclusion

From our European perspective, this Spiral Flow Graft, which restores the natural blood flow pattern, seems to be a strong tool to prevent stenosis at the AV graft venous anastomosis.

Back to top

8. Vascular Flow Technology: Another run-of-the-mill graft or a breakthrough technology? Experience and perspective from a US centre

Presented at the 35th Charing Cross International Symposium, London, UK 7 April 2013.
Hosam El Sayed, MD
Assistant Professor of Cardiovascular Surgery.
Methodist DeBakey Heart and Vascular Center, Houston, Texas, USA.

Patients and methods

From January, 2012 – February, 2013, 19 Spiral Flow™ AV Access Grafts (Vascular Flow Technologies Ltd) were placed in 18 patients. The mean age was 60 years (range 42-84 years); 37% (7) were male. Patients presented with the following comorbidities: six diabetes (32%); 17 hypertension (90%); five CAD (26%); four CHF (21%); five CVA (26%); six PVD (32%). Grafts were implanted as: 11 upper arm; two forearm; five femoral; and one chest wall.

Results

The mean follow-up was six months. There are 11 grafts (58%) still in use. Three grafts (15%) were removed for infection; two grafts thrombosed (10%); two grafts were ligated for severe steal (10%). There were two grafts with seroma (10%), but they remained functional. There were three wound complications (15%) and three arm swelling (15%). One graft was ligated (5%). There were no pseudoaneurysms. Two patients died with functioning grafts.
Overall patency results are as follows:

PrimaryPrimary AssistedSecondary
90% ± 9%90% ± 9%100% ± 0%

Conclusion

The early results are encouraging and tend to be better compared to standard straight ePTFE and heparin bonded grafts. This might be explained on the basis of improved haemodynamics created by the spiral laminar flow.

Back to top

Contact us today to find out how Spiral Laminar Flow can be integrated into your medical device technology.

Craig Dunlop, General Manager, Vascular Flow Technologies

General Manager

Craig Dunlop

Craig joined Vascular Flow in May 2014.  He comes with over 20 years of industry experience as a medical device engineer and manager.  Most recent positions held were at Vascutek Ltd as OEM Technical Manager and Development and Operations Manager for ePTFE Products.
Craig has a broad portfolio of successful new product developments and introductions within the Cardiovascular arena and has worked in both Europe and USA.
Craig was educated in Engineering and Operations Management at University of West of Scotland.

Kate Full at Vascular Flow Technologies

Financial Controller

Kate Full

Kate joined Vascular Flow Technologies from Braveheart Ventures where she was responsible for the evaluation and development of investment opportunities, focusing on financial diligence.
A Chartered Certified Accountant, she trained with PriceWaterhouseCoopers, qualifying in 2001 and continued in professional practice until joining Braveheart in 2004. Prior to embarking on her professional training Kate held a number of different accounting roles in a variety of industries including manufacturing, wholesale and leisure.

Angela Paterson, Head of QA & Regulatory Affairs, Vascular Flow Technologies

Head of Quality Assurance & Regulatory affairs

Angela Paterson

Angela joined Vascular Flow Technologies in April 2013 after spending the previous 14 years in Quality Assurance within a vascular graft manufacturing environment. Prior to this she spent three years in laboratories/ Biotech. Angela also has extensive experience working with regulatory agencies e.g. US FDA and Japanese PMDA.
Angela has a BSc in Applied Biosciences from Glasgow Caledonian University.

Jim Veitch, Head of Global Business Development, Vascular Flow Technologies

Head of Global Business Development

Jim Veitch

Jim has over 2 decades of experience in the Global Medical Device Business with a special interest in cardiovascular, LVAD, TAH and dialysis disease areas.
He works with numerous large and small device companies assisting in the successful promotion and commercialisation of their novel technologies.
This experience will help guide Vascular Flow through the next stages of their development and global expansion.

Graeme Houston, Non-Executive Director, Vascular Flow Technologies

Non-executive Medical Director

Graeme Houston

As Chair of Clinical Imaging and Intervention, University of Dundee, Professor Houston is a Consultant Cardiovascular Radiologist with an interest in endo-vascular therapies, and non-invasive vascular imaging using MRI and Ultrasound.
His medical training at St. Thomas’s Hospital in London followed first class honours science degrees in physics and pathology at Cambridge University.
His endo-vascular training was undertaken initially in Glasgow and further developed as a Fellow at the University of British Columbia, Vancouver, Canada. He is responsible for intellectual property, and as Medical Director is responsible for the clinical scientific research in the Spiral Laminar Flow™ field and stent technology.

Medical Device Development

Become an Investor

We are always interested to hear from potential investors with a view to accelerating our ambitious business plan and global reach or hearing from potential partners with a view to developing our Spiral Flow Technology in other areas.
If you want to find out more about investment opportunities in Vascular Flow Technologies, you can contact us directly on +44 1382 598 532.

Contact us
Medical device ISO class 7 clean room services

ISO Class 7 Clean Room

The Vascular Flow ISO Class 7 rooms provide air cleanliness of a maximum of 352,000 particles per cubic meter at 0.5μm and 60 to 90 air changes per hour. Vascular Flow use this cleanroom environment for medical device manufacturing and packaging processes.  Each client has dedicated cleanroom space and time.

Medical device Contract assembly services

Contract Assembly Services

Vascular Flow provide medical device contract assembly services to successfully deliver to our clients specifications. We assemble a wide range of medical devices for the world’s leading companies at low and medium volume manufacturing scales in our ISO Class 7 cleanroom environment.
Contact us today to find out how we can help you with your medical device contract assembly needs.

Medical device contract packaging services

Contract Packaging Services

Vascular Flow offer innovative packaging solutions for your medical device. Our ISO Class 7 cleanroom processes ensure 100% product or customer specific isolation to avoid contamination
We are specialists in low volume medical device contract packing projects for small and medium sized organisations. Our packaging solutions are optimised for process reliability to ensure consistent on time and on budget.
Contact us today to find out how we can help you with your medical device contract packaging needs.

injection moulding services

Injection Moulding Services

Our recognised know-how in plastic injection moulding enables Vascular Flow to offer our clients comprehensive and customised support, from the design of their devices right through to production, all in accordance with international regulatory requirements (ISO 13485-2016, ISO 15378 and QSR – 21 CFR part 820).

Medical device label printing services

Medical Device Label Printing Services

Vascular Flow Technologies are here to help get your products noticed. Our team is motivated to see your product succeed, whether you need a hundred labels or thousands of labels applied every day, we can meet your demand. Vascular Flow work with you to ensure only adhesives and topcoats that are compatible with your sterilization process are used.

Computer Aided design Services

Computer Aided Design Services

Vascular Flow utilises the latest Solidworks 3D CAD software to compliment its R&D offering.
Our highly skilled technical team utilises CAD to create and improve designs and reduce costs and is key to good communication of new products and documenting design history. We also have experience in parametric design processes and non-linear/complex curvature model creation.

Dummy Image

Finite Element Analysis

The team of experts at Vascular Flow can predict the structural performance of medical devices in conjunction with biological tissues using Finite Element Analysis (FEA).
Out team can couple together different physical effects using FEA, including heat and mass transfer, chemistry, fluid flow, electricity and solid mechanics.
We direct FEA to accelerate pace of innovation, decrease clinical trial costs, mitigate design risks and streamline the approval process.
Computational models are used to obtain information that are either too challenging or impossible to be obtained experimentally in pre-clinical or clinical studies.
To find out how Vascular Flow can assist you in your new development project, get in touch today.

In Vitro Flow Services (Bench Models)

In Vitro Flow Services (Bench Models)

In vitro tests with a bespoke, computer-controlled flow rig are used to verify the outcomes of medical devices in a controlled environment.
Vascular Flow has the capability of producing complex geometries and pulsatile flow to represent medical devices in blood vessels. We measure flow rates with pressure and flow sensors, and complex flow structures are measured with Doppler ultrasound. Blood viscosity and density are reproduced with blood mimic fluid.
In vitro flow & deployment (Cadaveric models)
Cadaveric models are the most realistic anatomical models to test vascular devices prior to pre-clinical trials.
Vascular Flow work in partnership with the University of Dundee and Tayside Health Board to test the deployment and flow outcome of vascular devices in a cadaveric anatomical model. The cadaveric bodies using the Thiel soft-fix method, which preserves them with life-like flexibility and tissue quality. Flow and anatomy are visualised with ultrasound, fluoroscopy, CT or MRI.

Project- Management and Delivery Services

Project Management & Delivery Services

Vascular Flow Technologies highly experienced team has a proven track record of taking products from initial design concept to market.
All projects delivered by Vascular Flow are ISO13485 compliant with Design History File generation.
We can work in collaborative partnerships or provide full turn-key solutions, offering flexible design, development & verification costing models.

Prototyping & Testing Services

Prototyping & Testing Services

To assist our clients to rapidly get their products to market, we offer a range of rapid prototyping services including injection moulding and plasma treatment.
We have vacuum chambers, furnaces with gas retort and laminar flow and extraction cabinets to allow us to safely and effectively work with your device design.
The product testing services we offer our clients are numerous and include:
· Burst testing
· Tensile testing
· Compression testing
· Suture retention testing
· Repeat puncture testing
· Permeability testing
· Package peel testing
· Dimensional testing
· Kink testing
· Flow testing
· Ultrasound doppler analysis
Our research team are highly adaptable and will work with you hand in hand to get the job done and move your project forward at pace.

Dummy Image

Regulatory Support

Vascular Flow Technologies has extensive Class IIa, IIb, III device submission experience with both CE and 510k submissions and approvals.
We can support full ISO 13485 design history and technical file generation & ISO14971 Risk Assessment.
Our experienced team can undertake full Clinical Literature Reviews on your behalf to support the early-stage feasibility processes.

Medical Device Development Services ISO Standard Performance Testing Services

ISO Standard Performance Testing Services

Vascular Flow has invested heavily in equipment and enhanced its laboratory to capitalise on the in house experience gained through previous projects and taking products to market.
This expertise allows us to offer product performance testing to ISO 7198 (Cardiovascular implants and extracorporeal systems) and ISO 10555 (Intravascular catheters).
Working within the constraints of our quality system ensures customers always receive comprehensive reports and test results.

Lucy Petersen, QA & RA Manager, Vascular Flow Technologies

QA & RA Manager

Lucy Petersen

Lucy joined Vascular Flow Technologies in January 2020 bringing with her knowledge and experience in quality and research and development gained in medical device manufacturing. Lucy also has experience in biocompatibility and toxicology as well as working with regulatory agencies e.g US FDA.
Lucy has a BSc in Chemistry from Aberdeen University.

Privacy Preferences

When you visit our website, it may store information through your browser from specific services, usually in the form of cookies. Here you can change your Privacy preferences. It is worth noting that blocking some types of cookies may impact your experience on our website and the services we are able to offer.

For performance and security reasons we use Cloudflare
required
Our website uses cookies, mainly from 3rd party services. Define your Privacy Preferences and/or agree to our use of cookies.